A broadband cavity-enhanced spectrometer for atmospheric trace gas measurements and Rayleigh scattering cross sections in the cyan region (470–540 nm)

dc.contributor.authorJordan, Nick
dc.contributor.authorYe, Connie Z.
dc.contributor.authorGhosh, Satyaki
dc.contributor.authorWashenfelder, Rebecca A.
dc.contributor.authorBrown, Steven S.
dc.contributor.authorOsthoff, Hans D.
dc.date.accessioned2019-03-01T23:02:13Z
dc.date.available2019-03-01T23:02:13Z
dc.date.issued2019-02-27
dc.description.abstractAn incoherent broadband cavity-enhanced absorption spectroscopy (IBBCEAS) instrument for quantification of atmospheric trace gases that absorb in the cyan region of the electromagnetic spectrum (470 to 540 nm), including NO2and I2, is described. The instrument uses a light-emitting diode coupled to a 1 m optical cavity consisting of a pair of mirrors in stable resonator configuration. Transmitted light is monitored using a grating spectrometer and charge-coupled device array detector. The average mirror reflectivity was determined from the N2/He and Ar/He ratios of scattering coefficients and was∼99.98 % at its maximum, yielding an effective optical path length of 6.3 km. Cross sections of N2, O2, air, Ar, CO2, and CH4 scattering and of O4 absorption were measured and agree with literature values within the measurement uncertainty. Trace gas mixing ratios were retrieved using the spectral fitting software DOA-SIS (DOAS intelligent system) from 480 to 535 nm. Under laboratory conditions, the 60 s, 1σ measurement precisions were ±124 and ±44 pptv for NO2and I2, respectively. The IBBCEAS instrument sampled ambient air in Ucluelet, BC,Canada, in July 2015. IBBCEAS retrievals agreed with in-dependent measurements of NO2by blue diode laser cavity ring-down spectroscopy (r2=0.975), but ambient I2concentrations were below the detection limit.en_US
dc.identifier.citationJordan, N., Ye, C. Z., Ghosh, S., Washenfelder, R. A., Brown, S. S. & Osthoff, H. D. (2019). A broadband cavity-enhanced spectrometer for atmospheric trace gas measurements and Rayleigh scattering cross sections in the cyan region (470–540 nm). "Atmospheric Measurement Techniques", 12, 1277-1293. http://dx.doi.org/10.5194/amt-12-1277-2019.en_US
dc.identifier.doihttp://dx.doi.org/10.5194/amt-12-1277-2019en_US
dc.identifier.urihttp://hdl.handle.net/1880/109922
dc.identifier.urihttps://doi.org/10.11575/PRISM/46146
dc.language.isoenen_US
dc.publisherCopernicusen_US
dc.publisher.departmentChemistryen_US
dc.publisher.facultyScienceen_US
dc.rightsUnless otherwise indicated, this material is protected by copyright and has been made available with authorization from the copyright owner. You may use this material in any way that is permitted by the Copyright Act or through licensing that has been assigned to the document. For uses that are not allowable under copyright legislation or licensing, you are required to seek permission.en_US
dc.rights.urihttps://creativecommons.org/licenses/by/4.0en_US
dc.titleA broadband cavity-enhanced spectrometer for atmospheric trace gas measurements and Rayleigh scattering cross sections in the cyan region (470–540 nm)en_US
dc.typeunknown
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Atmos_Meas_Tech_2019_12_1277.pdf
Size:
5.5 MB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.92 KB
Format:
Item-specific license agreed upon to submission
Description: